Home > LogTag TIC20

LogTag® TIC20 Disposable Temperature Indicator for Pharmaceutical Transit

The LogTag® TIC20 is an irreversible electronic temperature indicator for use in monitoring temperature & time statistics in pharmaceutical transport applications of up to 20 days duration.

The LogTag TIC20 can monitor and record statistics against four factory pre-set temperature limits for up to 20 days, displaying OK if environmental conditions
remain within limits. If temperatures exceed high temperature alarm limits or go below low temperature alarm limits, the LCD shows a REJECT alert, indicating the goods require further checking.

The LogTag TIC20 is factory pre-configured for one of a range of monitoring profiles. Special profiles are available on request, subject to conditions.

Statistical data can be downloaded from the TIC20 using the LogTag Indicator interface cradle and freely available LogTag Analyzer Software if required.

The TIC20 offers an accurate and cost effective solution for monitoring the cold chain in “last mile” distribution and other applications where full data logging capability is not required.

LogTag® TIC20 Temperature Indicator

LogTag Interface Software Kit
A combination of a single USB interface kit and a software disk. Plug and play with most systems.

Highlights

– Factory configured to monitor a range of products including standard vaccines
– Standard 60 minute start delay
– Clear indication of temperature history/alarm status at all times.
– Monitors for up to 20 days
– Push button Start and Stop. Push Review button to display daily statistics.
– Data can be downloaded via an interface cradle and LogTag Analyzer if required.
-.Robust and Accurate.
– Data collected is secure and cannot be manipulated.
– Fits easily into packages and can be mailed via letter rate.
– WHO PQS prequalified version available.

One year replaceable warranty against defects in materials and workmanship from the date of purchase. Excludes misuse or abuse of the product or as a result of unauthorized alteration or repairs.

LogTag® TIC20 – VACCINE SHIPPING TEMPERATURE INDICATORS

WHO PQS Versions – Prequalified under E06/06: PQS Code E006/021

Available in 4 standard configurations with labeling and instruction guides.

Also supplied: Backing cards in either English, French or Spanish and adhesive for attaching the recorder to the card (as shown here below)

Product Code: TIC20-W1

Product Code: TIC20-W2

Product Code: TIC20-WP

Product Code: TIC20-WR

Specifications
Model# TIC20
Operating temperature range -30°C to +60°C (-22°F to +140°F)
Ambient humidity range during transport and use 0% to 95%RH
Rated Temperature reading accuracy better than ±0.5°C (±1.5°F) for -10°C to +40°C (+14°F to +104°F).
better than ±1.0°C (±2.0°F) for -30°C to -10°C (-22°F to +14°F).
better than ±1.0°C (±2.0°F) for +40°C to +60°C (+104°F to +140°F).
Sensor Reaction Time (T90) T90 less than 7 minutes by method detailed in EN12830:1999.
Display Resolution 0.1°C from -30°C ~ +60°C (0.2°F from -22° ~ 140°F)
Capacity Minimum & maximum temperature for each of the 20 days. 1st activation of each alarm for each day including trigger time and duration of each excursion.
Sampling interval Factory set to 6 minute interval (i.e. 0.1hour)
Memory type Non volatile
Recordings 20 day statistics, temperature/time excursions and alarm trigger time and duration for each day.
Start Delay Factory set to 60 minutes
Environmental IEC 60529: IP64
Power source 3V Lithium-Manganese Dioxide, non replaceable
Battery life Minimum storage life of 18 months before ‘start’. Monitoring period: 20 days. Minimum accessibility (display) period of 6 months after ‘stop’
Size 73mm (H) x 54.5mm (W) x 8.6mm (T) (Volume <34cm³)
Weight 30 grams
Case Material Polycarbonate
Reader Interface LogTag TIC/TICT reader interface only
Calibration Certificate of Traceability and Calibration to ISO/IEC17025 available on request.
Warranty One year replaceable warranty against defects in materials and workmanship from the date of purchase. Excludes misuse or abuse of the product or as a result of unauthorized alteration or repairs.

Interface Cradle

The LogTag® TIC20/TICT reader interface is a special interface for reading the temperature indicator products and allows download of stored statistical and alarm trigger data for viewing & processing with LogTag Analyzer.

Standards Compliance

EC EMC directives (EN 61000-6-1:2005 & EN 61000-6-3:2006) Including electrostatic discharge as prescribed in EN 61000-4-2. Complies with FCC Part 15 Subparts A and B. Manufactured to meet the European RoHS Directive The RoHS Directive stands for “the restriction of the use of certain hazardous substances in electrical and electronic equipment”. This Directive bans the placing on the EU market of new electrical and electronic equipment containing more than agreed levels of lead, cadmium, mercury, hexavalent chromium, polybrominated biphenyl (PBB) and polybrominated diphenyl ether (PBDE) flame retardants.